SBE Water generation system ensures high quality water product by controlling microorganisms. SBE’s Water treatment units can Comply to strictest of regulatory as WHO, FDA, GAMP, cGMP, USP, EP, ChP etc.
| Unit | ChP | EP | USP | PW | WFI | CS | |
|---|---|---|---|---|---|---|---|
| Conductivity | µS/cm | <4.3@20°C | <4.3@25°C | <1.1@20°C | ≤0.5@20°C | ≤0.5@20°C | ≤1.3@20°C |
| TOC | mg/l | <0.5 | <0.5 | <0.5 | <0.5 | <50 | <0.5 |
| Microbe Limit | Cfu/ml | <100 | <100 | <100 | ≤50 | ≤10 | <100 |
| Endotoxin | EU/ml | <0.25 | <0.25 | <0.25 | - | <0.01 | <0.25 |